Waferpedia

Pfeiffer

ATH1603M

Deposition100mmPfeiffer ATH1603M family
Research Quality: 40% complete

The Pfeiffer ATH1603M is a magnetically levitated turbomolecular pump with a pumping speed of approximately 1,400–1,600 l/s for nitrogen. The ATH1603M achieves an ultimate pressure of up to 6×10⁻⁹ mbar and operates at 39,000 rpm. The ATH1603M is used as a secondary pump in vacuum systems, such as the Alliance-Concept EVA 451 thermal evaporator.[1][2]

PfeifferATH1603M
No photo available yet

Wafer size

100mm

Power

44 Vrms, 1,300 Hz, 7 Arms[3]

Vacuum

Turbo pump[2]

What it is

The Pfeiffer ATH1603M is a turbomolecular pump designed for generating high and ultrahigh vacuum.[2][1]

How it works

The ATH1603M achieves an ultimate pressure as low as approximately 6 × 10⁻⁹ mbar, depending on the system configuration.[1]

The pump is water-cooled and requires an external controller; it does not include an integrated drive electronics unit.[1][3]

Complementary dry pumps used in the same system include the Pfeiffer ACP series for primary pumping.[2]

What can it run?

Process applications and technology nodes documented for this tool.

  • Used in the pumping system of an evaporator for coating conductive layers
  • Associated system is used to prepare non-conductive samples for e-beam exposure and avoid charging issues

What do the numbers mean?

Power & electrical3

Motor parameters
44 Vrms, 1,300 Hz, 7 Arms[3]
Accurate?
Motor voltage
44 Vrms[1]
Accurate?
Motor frequency
1,300 Hz[1]
Accurate?

Vacuum & pumping4

Role
Turbo pump[2]
Accurate?
System location
Chamber secondary pump[2]
Accurate?
Pumping speed (N₂)
approx. 1,400–1,600 l/s[1]
Accurate?
Ultimate pressure
up to approx. 6 × 10⁻⁹ mbar[1]
Accurate?

Wafer handling1

Supported wafer sizes in associated system
100 mm wafers; 150 mm wafers[2]
Accurate?

Control & software2

Controller
External electronics required (e.g., MAGPOWER controller)[3]
Accurate?
Controller
External electronics required (without controller)[1]
Accurate?

Dimensions & facilities1

Cooling
Water-cooled[1]
Accurate?

Configuration & options10

Associated system
EVA 451 thermal evaporator[2]
Accurate?
Rated speed
39,000 rpm[1]
Accurate?
Bearing type
Magnetic bearing (MagLev)[1]
Accurate?
Inlet flange
DN250 ISO-F (with protective grille) / DN200 ISO-F[1]
Accurate?
Outlet flange
KF40 (NW40, ISO-KF40)[1]
Accurate?
Bearing
Magnetic bearing[1]
Accurate?
Inlet flange
DN250 ISO-F with protective grille[1]
Accurate?
Motor current
7 Arms[1]
Accurate?
Country of origin
France[4]
Accurate?
Inlet flange (alternate)
200 CF-F[5]
Accurate?

Vintage & configurations

Documented models & variants

DesignationGenerationVintageChangesSource
Y2662100———lagerwerk.com[1]
Y2522100———idealvac.com[5]
Y25231L0———idealvac.com[6]
TE5231B1———semiconductorspares.com[4]
ATH1603M (DN250)——Inlet flange DN250 ISO-F with protective grille, water-cooled, magnetic bearinglagerwerk.com[1]
ATH1603M (DN200) with MAGPOWER controller——Inlet flange DN200 ISO-F, includes MAGPOWER controller (type 120917 Ed.A), water cooling, magnetic bearinglagerwerk.com[3]
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • RoleTurbo pump[2]
  • System locationChamber secondary pump[2]
  • Pumping speed (N₂)approx. 1,400–1,600 l/s[1]
  • Ultimate pressureup to approx. 6 × 10⁻⁹ mbar[1]
  • Motor parameters44 Vrms, 1,300 Hz, 7 Arms[3]
  • Motor voltage44 Vrms[1]
  • Motor frequency1,300 Hz[1]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What controller options are compatible with the ATH1603M?

The pump can operate with a 115VAC or 220VAC controller, and some units do not have a fan.[5]

In which type of system is the ATH1603M used at EPFL?

At EPFL CMi, the ATH1603M is part of the Alliance-Concept EVA 451 thermal evaporator.[2]

Not publicly documented

The following facts about the ATH1603M are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the ATH1603M are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • The control-system platform and OS era of the ATH1603M are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the ATH1603M are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the ATH1603M is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

  • No publicly documented compatible parts, consumables, or accessories for the ATH1603M are on record.

    Answerable by: an OEM parts catalog or a service engineer

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (6)Every fact above is drawn from these public sources
  1. [1]lagerwerk.com — lagerwerk.comlagerwerk.com
  2. [2]epfl.ch — epfl.chepfl.ch
  3. [3]lagerwerk.com — lagerwerk.comlagerwerk.com
  4. [4]semiconductorspares.com — semiconductorspares.comsemiconductorspares.com
  5. [5]idealvac.com — idealvac.comidealvac.com
  6. [6]idealvac.com — idealvac.comidealvac.com
Was this page accurate?

Last updated Sep 30, 2026.

Compare with similar tools