370 A Programmable Curve Tracer

Provisional entry — research in progress

This page was generated automatically from cited public sources and hasn't completed the full research pass yet. Every specification shown carries its source; more detail is added as research completes.

370 A Programmable Curve Tracer — Inventory photo
Fig. 01370 A Programmable Curve TracerInventory photo[1]

Power

10 A (20 A pulsed)[2]

Optics

100 nA/div (1 nA resolution) to 2 A/div[2]

What it is

The Tektronix 370 A can source up to 2000 volts or 10 amps, with a maximum power of 220 watts. The instrument offers a measurement resolution of 1 nanoamp in current and 2 millivolts in voltage. The 370 A is fully programmable and features a 3.5-inch MS-DOS-compatible disk drive for saving and recalling setups. The instrument provides Kelvin sense measurements for accurate device characterization.[2]

How it works

The Tektronix 370 A includes features such as waveform comparison, envelope display, waveform averaging, and a dot cursor for precise measurement. The instrument supports Kelvin sense measurements to eliminate lead resistance errors. The 370 A is fully programmable and can store and recall setups from a 3.5-inch MS-DOS disk.[2][3]

Applications

The instrument is designed for high resolution DC parametric characterization of semiconductor devices, supporting both manual and automated testing.[2]

What do the numbers mean?

Power & electrical7

Maximum Peak Current (16 V range)
10 A (20 A pulsed)[2]
Accurate?
Maximum Peak Current (80 V range)
2 A (4 A pulsed)[2]
Accurate?
Maximum Peak Current (400 V range)
0.4 A (0.8 A pulsed)[2]
Accurate?
Maximum Peak Current (2000 V range)
0.05 A (0.1 A pulsed)[2]
Accurate?
Step Generator Modes
Stair step: DC, 80 µs pulse, 300 µs pulse[2]
Accurate?
Step Range (Voltage)
50 mV to 2 V in 1-2-5 sequence[2]
Accurate?
Measurement Resolution
1 nA current resolution, 2 mV voltage resolution[2]
Accurate?

Optics & imaging2

Collector Current Measurement Range
100 nA/div (1 nA resolution) to 2 A/div[2]
Accurate?
Emitter Current Measurement Range
1 nA/div (10 pA resolution) to 2 mA/div[2]
Accurate?

Performance2

Collector Current Accuracy
1.5% of cursor readout +0.05 div of setting (with dot cursor)[2]
Accurate?
Emitter Current Accuracy
1.5% of cursor readout +0.05 div of setting +1 nA[2]
Accurate?

Control & software1

Interface
IEEE Standard 488.1-1987, RS-232-C, Tektronix Standard Codes and Formats[2]
Accurate?

Configuration & options12

Built onto a cart[1]
Accurate?
Collector Supply Modes
AC, ±DC, ±Leakage, ±Rectified Sine[2]
Accurate?
Step Range (Current)
50 nA to 200 mA in 1-2-5 sequence[2]
Accurate?
Offset
Up to ±10X step amplitude[2]
Accurate?
Number of Steps
0 to 10[2]
Accurate?
Safety Certifications
UL1244, CSA231, HD401 S1 K475[2]
Accurate?
Storage
3.5-inch MS DOS-compatible disk, nonvolatile memory[2]
Accurate?
Test Fixture Adapters
A1001-A1005 (Kelvin sensing), 3-Pin adapters (non-Kelvin), A1023/A1024 surface mount adapters[2]
Accurate?
Accessories
K475 Instrument Cart[2]
Accurate?
Display Features
Waveform Comparison, Envelope Display, Waveform Averaging, Dot Cursor[2]
Accurate?
EMC Compliance
Directive 89/336/EEC, EN 55011 Class B (radiated and conducted), EN 60555-2[3]
Accurate?
Installation Category
CAT II (as defined in IEC 1010-1, Annex J)[3]
Accurate?
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • Maximum Peak Current (16 V range)10 A (20 A pulsed)[2]
  • Maximum Peak Current (80 V range)2 A (4 A pulsed)[2]
  • Maximum Peak Current (400 V range)0.4 A (0.8 A pulsed)[2]
  • Maximum Peak Current (2000 V range)0.05 A (0.1 A pulsed)[2]
  • Step Generator ModesStair step: DC, 80 µs pulse, 300 µs pulse[2]
  • Step Range (Voltage)50 mV to 2 V in 1-2-5 sequence[2]
  • Measurement Resolution1 nA current resolution, 2 mV voltage resolution[2]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What is the maximum voltage that the Tektronix 370 A can source?

The Tektronix 370 A can source up to 2000 volts.[2]

What is the maximum current that the Tektronix 370 A can source?

The Tektronix 370 A can source up to 10 amps.[2]

What is the maximum power that the Tektronix 370 A can provide?

The Tektronix 370 A can provide up to 220 watts.[2]

What measurement resolution does the Tektronix 370 A offer?

The Tektronix 370 A offers 1 nanoamp current measurement resolution and 2 millivolt voltage measurement resolution.[2]

Does the Tektronix 370 A support Kelvin sense measurements?

Yes, the Tektronix 370 A provides Kelvin sense measurements.[2][3]

Not publicly documented

The following facts about the 370 A Programmable Curve Tracer are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the 370 A Programmable Curve Tracer are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • No publicly documented variants, configuration options, or revision breakpoints of the 370 A Programmable Curve Tracer are on record.

    Answerable by: an OEM product catalog or an engineer who ordered or specified the tool

  • The control-system platform and OS era of the 370 A Programmable Curve Tracer are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the 370 A Programmable Curve Tracer are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the 370 A Programmable Curve Tracer is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

No research found yet — worked with this tool? Share what you know.