Waferpedia

Perkin Elmer

Optima 8000 ICP-OES

Precision OpticsPerkin Elmer Optima family
Research Quality: 50% complete

Provisional entry — research in progress

This page was generated automatically from cited public sources and hasn't completed the full research pass yet. Every specification shown carries its source; more detail is added as research completes.

Optima 8000 ICP-OES — Inventory photo
Fig. 01Optima 8000 ICP-OESInventory photo[1]

Power

40 MHz, free-running[4]

Optics

165–900 nm[2]

What it is

The Optima 8000 uses Flat Plate plasma technology, which replaces traditional helical induction coils with maintenance-free plates that reduce argon consumption. The instrument pairs a high-resolution echelle optical system with a simultaneous CCD detector operating across a 165–900 nm spectral range.[2]

How it works

The Optima 8000 incorporates a 40 MHz free-running RF plasma generator with an output power range of 750 to 1500 watts, controllable in 1-watt increments. The RF generator is located on the left side of the instrument. The system meets FCC certification requirements for RF emissions.[4]

The optical system of the Optima 8000 includes a double monochromator and a dual backside-illuminated CCD detector. Real-time Dynamic Wavelength Stabilization corrects for thermal drift. The instrument supports automatic dual viewing of the plasma, switching between axial and radial views to optimize sensitivity for trace and high-concentration elements.[4]

Applications

The system is particularly suited for EPA Method 6010D, which covers determination of metals in environmental samples. The Optima 8000's sensitivity and wide linear dynamic range allow simultaneous multi-element analysis in a single run, improving throughput for compliance monitoring.[2]

What do the numbers mean?

Power & electrical2

Rf Generator Frequency
40 MHz, free-running[4]
Accurate?
Power Output
750 to 1500 watts, computer-controllable in 1 watt increments[4]
Accurate?

Optics & imaging1

Spectral Range
165–900 nm[2]
Accurate?

Dimensions & facilities2

Dimensions (W x D x H)
52 x 32 x 30 inches (132 x 81 x 76 cm)[4]
Accurate?
Weight
310 lbs (141 kg)[4]
Accurate?

Configuration & options9

PolyScience WhisperCool Chiller[1]
Accurate?
Detection Limit
5 µg/L (ppb) to 100 mg/L (ppm)[3]
Accurate?
Output Power Stability
<0.1%[4]
Accurate?
Operating Temperature
15–35°C (59–95°F), max rate of change 2.8°C (5°F) per hour[4]
Accurate?
Optimum Room Temperature
20°C ±2°C[4]
Accurate?
Relative Humidity
20% to 80% without condensation; optimum 35%–50%[4]
Accurate?
Altitude
0–2000 m (sea level to 6562 feet)[4]
Accurate?
Sample Volume Required
2 mL for analysis[3]
Accurate?
Sample Introduction Volume (with Prep3)
2 mL[3]
Accurate?

Vintage & configurations

Documented models & variants

DesignationGenerationVintageChangesSource
Optima 8000 ICP-OES with Prep3——Includes Prep3.cest.nd.edu[3]
Optima 8000 ICP-OES with S10 Autosampler——Includes S10 autosampler.americanlaboratorytrading.com[4]
Optima 8000 ICP-OES with S10 Autosampler and PolyScience Chiller——Includes S10 autosampler and PolyScience WhisperCool refrigerated chiller.americanlaboratorytrading.com[4]
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • Rf Generator Frequency40 MHz, free-running[4]
  • Power Output750 to 1500 watts, computer-controllable in 1 watt increments[4]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What are the detection limits of the Optima 8000 ICP-OES?

The instrument can detect element concentrations as low as 5 µg/L (ppb) in solution and up to 100 mg/L (ppm).[3]

What is the wavelength range of the Optima 8000?

The instrument covers a spectral range from 165 to 900 nm.[2]

What autosampler is available for the Optima 8000?

The system can be configured with the Perkin Elmer S10 autosampler, which provides over 200 vessel positions and includes a built-in peristaltic pump for rinsing.[4]

What are the RF generator specifications of the Optima 8000?

The RF generator operates at 40 MHz with an output power range of 750 to 1500 watts, controllable in 1-watt increments, and power stability better than 0.1%.[4]

Not publicly documented

The following facts about the Optima 8000 ICP-OES are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the Optima 8000 ICP-OES are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • The control-system platform and OS era of the Optima 8000 ICP-OES are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the Optima 8000 ICP-OES are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the Optima 8000 ICP-OES is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

  • No publicly documented compatible parts, consumables, or accessories for the Optima 8000 ICP-OES are on record.

    Answerable by: an OEM parts catalog or a service engineer

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (5)Every fact above is drawn from these public sources
  1. [1]Inventory photo
  2. [2]amptechfl.com — amptechfl.comamptechfl.com
  3. [3]cest.nd.edu — cest.nd.educest.nd.edu
  4. [4]americanlaboratorytrading.com — americanlaboratorytrading.comamericanlaboratorytrading.com
  5. [5]PerkinElmer Optima 8000 ICP-OES Spectrometer &#8211; Spectralab Scientific Inc. — spectralabsci.comspectralabsci.com
Was this page accurate?

Last updated Sep 23, 2026.

Compare with similar tools

View all →