Empyrean

The Malvern Panalytical Empyrean is a multi-purpose X-ray diffractometer. The Empyrean combines X-ray diffraction, scattering, imaging, and X-ray absorption spectroscopy in a single instrument. The Empyrean uses MultiCore Optics to automatically configure the instrument for different measurements without manual intervention.[1]

Empyrean — Inventory photo
Fig. 01EmpyreanInventory photo[2]

Power

230 V[2]

Optics

0.026°[1]

What is it?

Applications

The Empyrean has been used in superconductivity research to characterize thin-film nickelate materials with X-ray diffraction using Cu Kα₁ radiation. In geoscience, it is employed for analysis of heterogeneous natural materials such as ores. The instrument is also applied to study battery materials, catalysts, and advanced alloys.[3][1]

What do the numbers mean?

Power & electrical4

Voltage
230 V[2]
Accurate?
Frequency
60 Hz[2]
Accurate?
X-ray generator maximum power
4 kW (max 60 kV, max 100 mA)[1]
Accurate?
Power supply
Single phase 210–230 V (±10% grid variations), 50/60 Hz[1]
Accurate?

Vacuum & pumping1

External cooling water supply
4–6 liters/min, pressure up to 8 bar, temperature 15 to 35 °C[1]
Accurate?

Optics & imaging3

Resolution (FWHM on LaB₆)
0.026°[1]
Accurate?
Detector types
PIXcel1D, PIXcel3D, GaliPIX3D; also 1Der, point detectors (proportional counter, scintillation detector)[1]
Accurate?
Source wavelengths
Cu Kα₁; also Mo radiation[3]
Accurate?

Dimensions & facilities2

Dimensions (W x D x H)
1400 × 1162 × 1947 mm[1]
Accurate?
Weight
1150 kg[1]
Accurate?

Configuration & options10

Power
8 kVa[2]
Accurate?
Max. Current
38 A rms[2]
Accurate?
Goniometer reproducibility
≤±0.0001°[1]
Accurate?
2θ linearity
±0.01°[1]
Accurate?
Maximum goniometer speed
15 deg/s[1]
Accurate?
Power consumption
Max. 4.6 kVA without controllers for optional equipment[1]
Accurate?
Compressed air supply
2–5 bar (0.2–0.5 MPa)[1]
Accurate?
Radiation level
Less than 1 microSievert per hour at 10 cm distance (Mo source @ 60 kV, 50 mA)[1]
Accurate?
Capabilities
X-ray diffraction (XRD), small- and wide-angle X-ray scattering (SAXS/WAXS), X-ray reflectivity, X-ray absorption spectroscopy (XAS), grazing incidence XRD, pair distribution function, texture mapping, residual stress measurement, computed tomography[1]
Accurate?
Temperature range for in situ stages
-200 °C to 1100 °C[4]
Accurate?
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • Voltage230 V[2]
  • Frequency60 Hz[2]
  • X-ray generator maximum power4 kW (max 60 kV, max 100 mA)[1]
  • Power supplySingle phase 210–230 V (±10% grid variations), 50/60 Hz[1]
  • External cooling water supply4–6 liters/min, pressure up to 8 bar, temperature 15 to 35 °C[1]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What measurement techniques does the Empyrean support?

The Empyrean combines X-ray diffraction, X-ray scattering, X-ray imaging, and X-ray absorption spectroscopy (XAS) in a single instrument. Specific techniques include SAXS, WAXS, grazing incidence XRD, pair distribution function analysis, residual strain measurement, microdiffraction, and computed tomography.[1][4]

What is the maximum temperature range for in situ sample stages on the Empyrean?

The Empyrean supports in situ sample stages that operate from -200 °C to 1100 °C (-328 °F to 2012 °F).[4]

What types of X-ray detectors are available for the Empyrean?

The Empyrean uses hybrid pixel detectors including the PIXcel3D and GaliPIX3D, which offer 0D, 1D, 2D, and 3D functionality with small pixel size, zero background, and high dynamic range.[1]

What is the X-ray generator power and maximum voltage/current of the Empyrean?

The Empyrean is equipped with a 4 kW X-ray generator with a maximum of 60 kV and 100 mA.[1]

What is the angular reproducibility of the Empyrean goniometer?

The goniometer reproducibility is ≤±0.0001°, and the smallest addressable increment is 0.0001°.[1]

Not publicly documented

The following facts about the Empyrean are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the Empyrean are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • No publicly documented variants, configuration options, or revision breakpoints of the Empyrean are on record.

    Answerable by: an OEM product catalog or an engineer who ordered or specified the tool

  • The control-system platform and OS era of the Empyrean are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the Empyrean are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the Empyrean is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (4)Every fact above is drawn from these public sources
  1. [1]malvernpanalytical.com — malvernpanalytical.commalvernpanalytical.com
  2. [2]Inventory photo
  3. [3]arxiv.org — arxiv.orgarxiv.org
  4. [4]adapt.mines.edu — adapt.mines.eduadapt.mines.edu
Was this page accurate?

Last updated Sep 25, 2026.

Compare with similar tools

View all →