Supra 55versusSupra 55 VP

Zeiss Supra 55, Zeiss Supra 55 VP, side by side. Every value shown is drawn from the cited encyclopedia entries.

Side-by-side comparison of selected equipment models
Attribute
OEMZeissZeiss
CategoryMetrology & InspectionMetrology & Inspection
Wafer size——
Process node——
Introduced——
Production run——
Lifecycle——
Lifecycle milestones——
Control system——
Generation——
FamilyZeiss Supra 55Zeiss Supra 55
Cited variants
  • Supra 55 VP[1]
  • Supra 55 FESEM[2]
—
Specifications
ConfigurationDetectors:[3]Variable pressure[6]
Electron Source TypeSchottky thermal field emission source (FESEM)[2]—
Accelerating Voltage0.2 to 30 kV[4]—
Accelerating Voltage (VP variant)0.1 to 30 kV[5]—
Imaging Resolution (High Vacuum, 15 kV)1.0 nm[5]—
Imaging Resolution (High Vacuum, 1 kV)1.7 nm[5]—
Imaging Resolution (High Vacuum, 0.2 kV)3.5 nm[5]—
Imaging Resolution (VP mode, 30 kV)2.0 nm[5]—
Resolution (general range)2 nm[2]—
Magnification Range12 - 900,000x[5]—
Probe Current4 pA to 10 nA (20 nA optional)[5]—
Variable Pressure Range2 to 133 Pa, adjustable in steps of 1 Pa[5]—
DetectorsInLens SE, E-T (SE2), AsB (backscattered), CL (cathodoluminescence), STEM, EDS, EBSD[3]In-lens detector, Everhart-Thornley detector, secondary detector, VP-SE detector, and 4-quadrant backscattered electron detector[7]
Sample Stage5-Axes (X, Y, Z, R, T) Motorized Cartesian[5]—
Operating SystemWindows 7 32-bit with Smartsem v5.05[3]—
Load-Lock SystemFJELD Airlock[3]—
PumpsAdixen rough pump (airlock), Edwards RV12 oil rough pump (chamber)[3]—
ChillerATC Chiller[3]—
ModelSupra 55[3]—
Equipment typeScanning Electron Microscope (SEM)[3]—
Source typeFESEM[2]—
Electron sourceSchottky thermal field emission source[2]—
Electron image resolution1.0 nm (@ 15 kV), 1.7 nm (@ 1 kV), 3.5 nm (@ 0.2 kV), 2.0 nm (@ 30 kV in VP mode)[5]1.0 nm at 15 kV; 1.7 nm at 1 kV; 3.5 nm at 0.2 kV; 2.0 nm at 30 kV in VP mode[5]
Magnification12 - 900,000x[5]12 to 900,000x[5]
Variable pressure range2 - 133 Pa, adjustable in steps of 1 Pa[5]2 to 133 Pa[5]
Acceleration voltage0.1 - 30 kV[5]—
Probe current4 pA - 10 nA (20 nA optional)[5]4 pA to 10 nA[5]
AccessoryFJELD Airlock[3]—
Windows OS—W10[6]
Software—SmartSEM (v6)[6]
Source Type—TFE[6]
Primary pump—turbo[6]
Instrument type—Field emission scanning electron microscope (FE-SEM)[5]
Operating modes—High vacuum and variable pressure[5]
Field emission cathode—Present[7]
Accelerating voltage range—100 V to 30 kV[7]
Sample stage—5-axes motorized Cartesian sample stage[5]
Probe current optional upper limit—20 nA optional[5]
Accessories—EDS detector, EBSD[5]
Control system OS—Windows 10[6]
Secondary electron detector—Everhart-Thornley secondary electron detector[6]
In-lens detector—Present[6]

Plan your fab

Building a Metrology & Inspection process? Get a complete equipment list.

Open the fab equipment planner →
Sources (8)Every fact above is drawn from these public sources
  1. [1]bu.edubu.edu
  2. [2]cmdis.rpi.educmdis.rpi.edu
  3. [3]inventory listing
  4. [4]weizmann.ac.ilweizmann.ac.il
  5. [5]binghamton.edubinghamton.edu
  6. [6]inventory listing
  7. [7]tuhh.detuhh.de
  8. [8]bu.edubu.edu